Baseline Hand Dynamometer Setup & Recording by Model
Start a baseline hand dynamometer test by identifying the model and readout. A mechanical dial, digital LCD display, and transducer output need different controls. Then record each trial with its units and testing conditions so the next assessment has a useful point of comparison.
Our baseline hydraulic hand dynamometers now include the models below. Use the product page’s model selector to match the reference on your instrument. For setup, the important distinction is how you obtain and clear the reading.
Match the model to its recording workflow.
| Model and reference | Readout and capacity | What changes in your workflow |
|---|---|---|
| Standard · 12-0240 | Mechanical dial · 200 lb | Read the retained peak; use the dial reset procedure. |
| LiTE · 12-0241 | Mechanical dial · 200 lb | Use the same dial workflow; identify this model in the record. |
| HD · 12-0221 | Mechanical dial · 200 lb | Record the instrument identity rather than treating all 200 lb models as the same device. |
| HiRes · 12-0243 | Larger mechanical dial · 200 lb | Use the scale on this dial; a larger face does not increase its capacity. |
| HiRes ER · 12-0246 | Larger mechanical dial · 300 lb | Check the extended-range scale before recording the peak. |
| Digital LCD ER · 12-0247 | LCD · 300 lb | Distinguish current force, retained MAX, zeroing, and MAX clearing. |
| Reverse gauge · 12-0271 | Participant-facing dial · 200 lb | Record whether the participant saw feedback and whether the task was testing or target-force practice. |
| Transducer · 12-0021 | Dial plus analog output · 300 lb | Identify whether the reported value came from the dial or an external recording system. |
| Transducer · 12-0248 | Analog output only · 300 lb | Use a compatible receiving system; there is no built-in gauge to read. |
Capacity is a measurement limit, not a target score. A 300 lb range does not prescribe a harder squeeze. A larger dial or numeric display also does not, by itself, establish better clinical outcomes.
Start with the testing plan. A treating clinician should determine whether maximum-effort grip testing is appropriate, especially after injury or surgery. Confirm the protocol before testing; stop if the test causes pain.
Set the handle and define the test before measuring.
Select the prescribed handle position and check that it is secure. For repeated assessments, keep the instrument and handling setting consistent whenever the clinical plan allows. Record a necessary change, including a switch from one baseline model to another.
The hand-evaluation instructions (PDF) describe an arm position with the upper arm beside the body, shoulder in neutral rotation, elbow bent to 90 degrees, and forearm and wrist neutral. Their grip-test description uses three trials and an average. Record your protocol’s posture, hand order, effort duration, rest interval, instructions, and summary method.
Handle position 2 is protocol-specific. It was used with a Jamar dynamometer in Mathiowetz and colleagues’ 1985 adult reference study. That paper specified its own positioning, too. This historical example does not make one grip setting suitable for every hand or every assessment.
Mechanical dial models: read and reset the peak pointer.
This workflow applies to the Standard, LiTE, HD, HiRes, and HiRes ER dial models. The reverse-gauge model also needs the feedback distinction explained below. Refer to FEI’s dial-model instructions (PDF) for the labeled controls.
- Inspect the unloaded instrument. Check for damage or leakage and confirm that the black indicator needle rests in the zero range.
- Clear the previous peak. Gently turn the peak-reset knob counterclockwise until the red maximum pointer meets the black pointer at zero, following the hand-evaluation instructions. The separate zero-adjustment pin is not the routine reset knob.
- Perform the planned squeeze. Read the retained maximum on the selected lb or kg scale. View the dial directly and use its marked divisions.
- Write down the trial before resetting. Include units and any interruption. Use the planned rest period before another attempt.
The red pointer staying at a high value after release is its peak-hold function. It does not calculate the mean of several trials.
Digital LCD 12-0247: separate ZERO, MAX, and MAX CLR
The digital LCD model uses the MENU and ON/ENTER buttons. Press ON/ENTER to switch power on or off. For setup, follow the press-and-hold flowchart in FEI’s 12-0247 Digital LCD instructions (PDF); a brief press and a hold have different functions.
| Menu function | What it does | Recording implication |
|---|---|---|
| ZERO UNIT | Sets the unloaded zero offset. | Zeroing and clearing a previous maximum are separate operations. |
| MAX | Displays the retained maximum. | Use it when the protocol calls for a trial peak. |
| MAX CLR | Clears the stored maximum through the confirmation sequence. | Save the previous trial first; confirm it is cleared before a new trial. |
| LB/KG selection | Selects the displayed unit. | Check the selected unit when transcribing the result. |
Check the unloaded zero and unit, clear the previous MAX, perform the agreed trial, and record its MAX before clearing again. The current-force display and retained maximum are different readings; a current value after release is not the trial peak. Write the result into the clinical record rather than relying on the value remaining on the screen.
Reverse gauge 12-0271: distinguish a test from feedback practice
This model places the dial toward the person squeezing, providing visual force feedback. “Reverse” describes the gauge orientation, not a finger-extension test.
FEI also illustrates using the maximum indicator as a target marker for a prescribed feedback exercise. If that is the task, record the target and practice instructions as such; do not enter the preset target as a measured maximum. For a peak-strength assessment, use the appropriate reset-and-test procedure.
Document whether the participant could see the readout, and keep that condition consistent for comparison. FEI’s feedback-model illustration (PDF) shows the intended arrangement; its sample exercise is not an individual treatment prescription.
Transducer models: identify where the result comes from.
12-0021 offers a dial and an analog output. 12-0248 provides analog output without a built-in gauge. With the output-only model, a compatible receiving system is needed to obtain a displayed or recorded measurement.
Before external capture, have the complete measurement system configured and checked under its documentation. Confirm the interface and signal-to-force conversion; do not label a raw voltage value as pounds or kilograms. Document the receiving system, software, and the rule used to extract a trial peak.
For 12-0021, state whether each result came from the onboard dial or the external system. Keep the source consistent across visits where possible. An analog-output specification alone does not confirm compatibility with a particular app, USB interface, or wireless recorder.
Keep a trial peak, the best trial, and the mean separate.
A trial peak is the highest force within one squeeze. The best trial is the largest result in a series. A mean is the sum of the trial results divided by their count. Decide which statistic your protocol requires before collecting results.
| Recorded item | Value | Meaning |
|---|---|---|
| Trial 1 / Trial 2 / Trial 3 | 38 / 42 / 40 lb | Three separately recorded peaks |
| Mean of three | 40 lb | (38 + 42 + 40) ÷ 3 |
| Best trial | 42 lb | The largest trial, not their average |
For a three-trial mean protocol, report 40 lb, the mean of three, and retain the raw values. Do not substitute the best trial at the next visit. A skipped or interrupted test is not a measured zero: record the reason and follow the protocol for incomplete testing. These example values are not normal-strength thresholds or recovery targets.
Grip-strength recording sheet
For a three-trial mean protocol. Complete the setup fields before testing. Use a separate row for each hand; mark a test as “not performed” when appropriate.
Date/time: Record ID: Examiner: Hand dominance:
Model/reference: Serial/asset ID: Units: □ lb □ kg Protocol name/version:
Readout source: □ dial peak □ LCD MAX □ external-system peak The participant could see the readout: □ yes □ no
Seated/arm/wrist position: Hand order: Effort duration: Rest interval: Instruction cue:
| Hand | Handle position | Trial 1 | Trial 2 | Trial 3 | Mean of 3 |
|---|---|---|---|---|---|
| Right | |||||
| Left |
External system/software and peak-processing rule, if used:
Symptoms, interrupted trials, or changes from the previous assessment:
Before comparing visits, confirm the same device, readout source, handle position, units, posture, feedback conditions, timing, and summary method. Document any necessary change.
Copy the sheet into your documentation template or print it. Ask about hand dominance and identify right and left separately. For a feedback exercise, use a separate exercise record so a target-force task is not mixed into a maximum-strength test series.
Resolve a questionable reading before interpreting change.
A dial needle is outside the unloaded zero range: resetting the maximum pointer is not a repair. Follow the manufacturer’s zero/service instructions before using the result. A zero check alone does not verify accuracy throughout the range.
A digital maximum seems unchanged: check whether MAX was cleared before the new trial and whether you are viewing current force or the retained maximum. Do not relabel an old peak as a new measurement.
The instrument is damaged or leaking: stop using it for assessment and arrange evaluation. Clean grip surfaces between patients and store the device in its protective case.
A later score is different: first check the model, readout source, and units; then handle setting, posture, feedback, timing, and calculation. A clinician should interpret the result with symptoms and function. Moving from a dial to an LCD is not, by itself, evidence of improvement or decline.
Common questions about model choice and test records
Does a digital display mean the dynamometer is no longer hydraulic?
No. The 12-0247 remains a hydraulic hand dynamometer; the LCD changes how you read and control the display. “Digital” and “hydraulic” describe different features of this model.
Does clearing MAX recalibrate the device?
No. It clears the retained maximum. Zeroing, clearing a peak, and checking measurement accuracy are different tasks. Use the procedure for your exact instrument and your facility’s equipment-quality process.
Can I compare any model’s result with an online age chart?
Check the original reference population and method first, including instrument compatibility, hand, units, grip setting, posture, and reported statistic. A matching unit label alone does not establish an equivalent test. An age-group average is not a diagnosis or an individual treatment goal.
Continue from measurement to an established exercise plan
A measured grip result does not directly select a resistance for CanDo Digi-Flex or a firmness for TheraPutty Standard. Our therapy putty, Digi-Flex, and Digi-Extend comparison explain exercise-format choices; our therapy putty exercise guide covers prescribed movements. Browse our hand-therapy equipment collection for current assessment and exercise products.
Copyright © Balego Inc.









