Salt-Free vs. Saline Electrode Gel: What’s the Difference?

by Service, Training and Support·September 12, 2026

Salt-Free vs. Saline Electrode Gel: What’s the Difference?

Does an electrode gel need salt to conduct electricity?

No.

Both salt-free and saline electrode gels can provide an electrically conductive interface between intact skin and compatible medical electrodes. The important difference is how the formulation achieves and maintains that interface—and which procedures the manufacturer specifically recommends it for.

Parker Laboratories provides a useful direct comparison:

Spectra 360 is a salt-free, chloride-free conductive electrode gel designed to wet the skin, reduce skin resistance, and remain useful during prolonged electrode contact.

SignaGel is a multipurpose saline electrolyte gel positioned around high conductivity and broad electromedical use, particularly ECG, EMG, biofeedback, and compatible defibrillation procedures.

The quickest selection rule is

Choose Spectra 360 when salt-free/chloride-free formulation, longer contact, TENS, monitoring, pediatrics, or conductive rubber/carbon electrodes are priorities. Choose SignaGel when the workflow specifically calls for a multipurpose saline electrode gel, particularly for ECG, EMG, biofeedback, or compatible defibrillation.

Shop Spectra 360 Electrode Gel.

Shop SignaGel Electrode Gel.

Salt-Free vs. Saline Electrode Gel: Quick Comparison

FeatureSpectra 360 — Salt-FreeSignaGel—Saline
Product categoryConductive electrode gelConductive electrode gel
Salt-freeYes.No— it as saline
Chloride-freeYes.Not Parker's distinguishing claim
How Parker describes conductivityWets skin and helps reduce skin resistanceHighly conductive multipurpose electrolyte gel
Long-term applicationsSpecifically highlightedGeneral electromedical use
STAY-WET formulationYes.No equivalent claim
ECG/EKGYes.Yes.
TENSSpecifically recommendedUse when device/electrode instructions permit.
Patient monitoringSpecifically recommendedProcedure/equipment dependent
Pediatric useSpecifically listedNot a primary current Parker claim
Conductive rubber/carbon electrodesPreferred by ParkerUse when compatible.
EMGFollow equipment instructions.Specifically recommended
BiofeedbackFollow equipment instructions.Specifically recommended
DefibrillationNoYes, when equipment permits separate gel
Water-solubleYes.Yes.
Non-grittyYes.Yes.
BacteriostaticYes.Yes.
SterileNoNo
SkinIntact, unbroken skinIntact, unbroken skin

Spectra 360 IFU specifically identifies it as nonsterile, salt-free, chloride-free, and electrically conductive; says its salt-free characteristics make it particularly suitable for long-term applications; and lists ECG, TENS, and monitoring while excluding defibrillation. Parker separately describes SignaGel as its ideal saline electrode gel and recommends it for ECG, defibrillation, biofeedback, and EMG.

What Is Saline Electrode Gel?

A saline electrode gel is an electrically conductive contact medium that uses an electrolyte-containing formulation.

Dissolved electrolytes provide mobile ions that support electrical conduction through the gel between the electrode and skin.

SignaGel as a:

“highly conductive, multi-purpose electrolyte” meeting the characteristics of an ideal saline electrode gel.

In practical use, a saline gel can help:

  • create a continuous interface between electrode and skin;
  • fill microscopic gaps;
  • improve electrical contact;
  • support acquisition of small physiological signals; and
  • Provide a consistent conductive medium during compatible electromedical procedures.

Saline does not mean that the product is simply ordinary salt water.

SignaGel is a purpose-formulated medical electrode gel with its own viscosity, bacteriostatic properties, skin-contact instructions, and device indications.

View SignaGel Electrode Gel.

What Is Salt-Free Electrode Gel?

A salt-free electrode gel provides electrical contact without relying on a conventional salt-containing formulation.

Spectra 360 is a primary example.

Its current IFU describes the product as

  • salt-free;
  • chloride-free;
  • electrically conductive;
  • non-gritty;
  • bacteriostatic;
  • suitable for prolonged use without reapplication; and
  • particularly suited to long-term applications.

Parker explains its operating principle simply:

Spectra 360 wets the skin, thereby reducing skin resistance.

That makes an important point:

An electrode gel does not have to contain added salt to function as a conductive contact medium.

Does Salt-Free Mean Less Conductive?

Not necessarily.

This is probably the most important misconception for this topic.

Saline solutions conduct electricity effectively because dissolved ions can carry charge.

But that does not mean:

More salt = better electrode gel.

or

Salt-free = poor conductivity.

An electrode procedure involves much more than conductivity of the gel by itself.

Performance also depends on:

  • skin resistance;
  • electrode construction;
  • electrode surface area;
  • contact pressure;
  • hydration;
  • gel coverage;
  • duration;
  • device output characteristics; and
  • the electrical requirements of the procedure.

Both Spectra 360 and SignaGel are electrically conductive products. However, Parker does not publish a simple apples-to-apples conductivity ranking showing that one universally provides better electrical performance than the other.

Therefore, avoid choosing between them by asking:

“Which one conducts the most electricity?”

The better question is

“Which formulation and manufacturer indication match my electrode and procedure?”

Why Make an Electrode Gel Without Salt?

Spectra 360's salt-free characteristics are advantageous for long-term use, including avoiding sodium-ion transfer and helping preserve electrode performance. Its STAY-WET formula is intended to remain moist during prolonged applications.

That makes salt-free formulation especially relevant when treatment or monitoring involves extended skin contact.

Situations favoring salt-free Spectra 360

  • patient monitoring;
  • longer electrode-contact periods;
  • TENS with compatible reusable electrodes;
  • pediatric procedures;
  • conductive carbon electrodes;
  • conductive rubber electrodes; and
  • situations where salt-free/chloride-free formulation is specifically desired.

This does not establish that every saline electrode gel causes irritation or electrode damage.

It means deliberately designed and marketed Spectra 360 around a different use profile.

Why Is Chloride-Free Important?

Spectra 360 is not merely labeled salt-free. Parker also specifically identifies it as chloride-free.

Parker connects its no-sodium-ion-transfer design with long-term use and preservation of electrode performance.

The appropriate conclusion is therefore product-specific:

Spectra 360 provides a conductive interface without salt or chloride and is intentionally positioned for prolonged contact and reusable conductive electrodes.

It would be too broad to conclude that every chloride-containing or saline gel damages equipment.

The electrode instructions remain the deciding authority.

Is Salt-Free Electrode Gel Better for Skin?

Not universally.

Parker describes Spectra 360 as non-irritating and positions its salt-free formulation for sensitive and longer-term applications.

But salt-free does not mean allergy-free or irritation-proof.

Spectra 360's current IFU specifically warns:

Do not use it in patients with a known history of sensitivity to parabens.

It is also

  • nonsterile;
  • for external use;
  • limited to intact, unbroken skin; and
  • subject to the electrode/device instructions.

The better language is therefore

Spectra 360 is designed for prolonged and sensitivity-conscious applications—not that it is universally safer for every patient.

Spectra 360 vs. SignaGel: Which Should You Choose?

The most useful distinction is the actual procedure.

Choose Spectra 360 when:

  • You specifically want a salt-free electrode gel
  • A chloride-free formulation is preferred;
  • Treatment or monitoring may involve longer contact;
  • You are using compatible TENS equipment;
  • You use conductive carbon or rubber electrodes
  • Monitoring is a primary application;
  • pediatric use is part of the workflow; or
  • STAY-WET formulation fits the procedure.

Spectra 360 is the preferred product for conductive rubber and carbon electrodes.

Shop Spectra 360

Choose SignaGel when:

  • The equipment instructions call for saline electrode gel;
  • ECG/EKG is the primary procedure;
  • EMG is being performed;
  • Biofeedback is being performed;
  • A multipurpose saline gel is preferred; or
  • A compatible defibrillation workflow requires separately applied electrode gel.

SignaGel for ECG, defibrillation, biofeedback, and EMG.

Shop SignaGel

Which Is Better for ECG: Salt-Free or Saline Gel?

Either type can be appropriate depending on the ECG electrode and equipment.

ECG in the manufacturer-listed applications for both Spectra 360 and SignaGel.

The selection becomes clearer when other priorities are added.

ECG situationBetter starting point
Routine ECG calling for a general saline electrode gelSignaGel
Longer patient monitoringSpectra 360
Salt-free/chloride-free formulation specifically desiredSpectra 360
Pediatric ECG/contact applicationSpectra 360 is specifically positioned
ECG-plus-compatible defibrillation useSignaGel
Pre-gelled disposable ECG electrodeSeparate gel is often unnecessary.

Do not add either gel beneath a self-contained pre-gelled ECG electrode unless that electrode's instructions call for additional conductive medium.

Which Is Better for Long-Term Monitoring?

Spectra 360 has the clearer manufacturer positioning.

Parker specifically states that:

  • Its non-gritty formulation permits prolonged use without reapplication
  • Salt-free characteristics make it particularly suitable for long-term applications; and
  • Monitoring is one of its listed procedures.

Its STAY-WET formulation is also specifically marketed around maintaining moisture during prolonged contact.

That makes Spectra 360 the stronger first comparison when contact duration is a major purchasing factor.

This does not mean SignaGel must be removed after a specific number of minutes. Parker simply gives Spectra 360 more explicit long-duration positioning.

Which Is Better for TENS?

For a setup requiring separate electrode gel, Spectra 360 has the clearer Parker recommendation.

Parker specifically lists TENS and identifies Spectra 360 as its preferred gel for conductive rubber and carbon electrodes.

This is particularly relevant to reusable:

  • carbon rubber electrodes;
  • conductive rubber electrodes; and
  • other compatible non-gelled TENS electrodes.

For that specific workflow, we already have a deeper application guide.

What Conductive Gel Should You Use With Carbon Rubber TENS Electrodes?

What about modern self-adhesive TENS pads?

Most self-adhesive TENS/NMES electrodes already incorporate hydrogel.

They ordinarily do not need either Spectra 360 or SignaGel added beneath them unless the electrode manufacturer explicitly permits it.

Is SignaGel Suitable for TENS?

Do not interpret this as “SignaGel cannot conduct TENS.”

It is an electrically conductive medical gel.

The issue is manufacturer positioning.

Parker specifically highlights SignaGel for:

  • ECG;
  • EMG;
  • biofeedback; and
  • defibrillation.

By comparison, Spectra 360 for TENS and carbon/rubber electrodes.

Therefore, if the question is simply

“Which Parker electrode gel should I select for reusable carbon-rubber TENS electrodes?”

Spectra 360 is the more directly supported choice.

If a TENS or electrode manual specifically calls for another compatible gel, follow those instructions.

Which Is Better for EMG and Biofeedback?

SignaGel has the stronger direct manufacturer indication.

SignaGel for both EMG and biofeedback.

Spectra 360 is broadly recommended for electromedical procedures other than defibrillation, but its specifically highlighted uses are

  • ECG;
  • TENS;
  • monitoring; and
  • pediatrics.

For EMG or biofeedback, SignaGel is therefore the more straightforward Parker starting point unless the electrode/equipment manufacturer directs otherwise.

Which Gel Can Be Used for Defibrillation?

This is the clearest situation where the two products should not be treated as interchangeable.

SignaGel

Parker includes defibrillation among SignaGel's recommended procedures.

Spectra 360

Parker expressly states:

Spectra 360 is not indicated for defibrillation.

Defibrillation questionAnswer
Can SignaGel be used?Yes, when the defibrillator/electrode IFU permits separate gel
Can Spectra 360 be substituted because it is conductive?No
Does every modern defibrillator need separate gel?No
Should gel be added to pre-gelled multifunction pads?Only if the pad/device instructions specifically say to

Many contemporary defibrillation electrodes already contain their own conductive interface.

Always follow the defibrillator and electrode instructions.

Does Saline Gel Give a Stronger TENS Sensation?

There is no reliable basis for predicting treatment intensity simply from saline vs. salt-free formulation.

How strong electrical stimulation feels also depends on:

  • device output;
  • waveform;
  • pulse width;
  • electrode area;
  • electrode placement;
  • skin contact;
  • tissue anatomy; and
  • individual sensory response.

Likewise, an ECG tracing depends on the complete electrode-skin-device system rather than one formulation label.

Do not use “stronger sensation” as a test of whether one gel is better.

The goal is appropriate and consistent electrical contact.

Does More Salt Mean Better Conductivity?

Not as a clinical purchasing rule.

In basic chemistry, increasing mobile ions can affect the conductivity of a liquid.

But medical electrode gel is not simply a chemistry demonstration.

Increasing salt content without regard to:

  • formulation;
  • skin compatibility;
  • moisture retention;
  • electrode materials;
  • device design; and
  • intended procedure

does not establish a better electrode interface.

This is why Parker sells both a highly conductive saline product and a salt-free conductive product.

Each is engineered for a different clinical use profile.

Never add table salt, homemade saline, or another ingredient to a medical electrode gel.

How Does Spectra 360 Conduct Without Salt?

Parker's current explanation is intentionally straightforward:

It wets the skin and thereby reduces skin resistance.

Skin itself is a significant part of the electrical interface.

Dry surface contact can increase resistance and produce inconsistent coupling.

By maintaining a moist, continuous contact layer, Spectra 360 helps create the electrical interface needed by compatible electrodes without relying on Parker's traditional saline-gel approach.

The precise electrical behavior still depends on the entire electrode/device system.

Does Salt-Free Mean Sodium-Free?

For Spectra 360 specifically, Parker states that its salt-free formulation provides no sodium ion transfer.

That is a characteristic of Spectra 360.

Is Chloride-Free the Same as Salt-Free?

Not as a universal chemistry definition.

“Salt” describes a broad class of ionic compounds, while chloride refers specifically to compounds containing chloride ions.

For Spectra 360, Parker explicitly claims both:

salt-free

and

chloride-free.

Those two claims should therefore be repeated together when describing this specific product rather than treating “chloride-free” as an automatic synonym for every salt-free gel.

Does Saline Electrode Gel Corrode Electrodes?

Do not make that generalization.

Parker markets Spectra 360's no-sodium-ion-transfer formulation as useful for preserving electrode performance during long-term use, but that does not establish that SignaGel or every saline medical gel will corrode every compatible electrode.

The material matters.

A disposable ECG electrode, a reusable carbon-rubber electrode, and a reusable metal electrode can have completely different contact requirements.

Follow the electrode manufacturer's instructions.

Carbon Rubber Electrodes: Salt-Free Has a Clear Advantage in Product Selection

For reusable conductive rubber or carbon electrodes, Spectra 360 is not merely a theoretical alternative.

Parker specifically calls it the preferred product for that electrode type.

That gives clinicians a very simple purchasing decision:

Electrode typeSeparate gel approach
Conductive rubber/carbon electrode requiring gelSpectra 360 is Parker's preferred option.
Self-adhesive hydrogel electrodeSeparate gel is usually unnecessary
ECG electrode specifically calling for saline gelSignaGel may fit.
Electrode needing adhesive conductive mediumConsider Tensive when permitted.
Non-gelled neurodiagnostic cup electrodeUse the specified neurodiagnostic paste.

Read our carbon-rubber TENS electrode gel guide.

Salt-Free Gel vs. Adhesive Gel: They Solve Different Problems

Salt-free is a formulation characteristic.

Adhesive is a mechanical property.

Spectra 360 is not an adhesive gel.

If an electrode needs both conductivity and temporary hold, Tensive Conductive Adhesive Gel is a different product category. It is for compatible TENS, EMG, EEG, and other electromedical applications where adhesive conductive gel is appropriate; it is not indicated for defibrillation.

Compare Tensive Conductive Adhesive Gel

Do not choose Tensive merely because the electrode gel keeps moving. Confirm that an adhesive contact medium is permitted for that electrode first.

Salt-Free Electrode Gel Is Not Skin-Prep Gel

Another common category mistake is confusing conductive gel with skin preparation.

Spectra 360 remains between the electrode and skin during the procedure.

NuPrep performs a different job: it is a mildly abrasive skin-preparation gel intended to help lower electrode-site impedance before compatible monitoring electrodes are applied. Balego's current product documentation specifically warns against using NuPrep with current-inducing electrodes.

Therefore:

Spectra 360 / SignaGel → conductive interface

NuPrep → skin preparation

They are not substitutes.

Salt-Free Electrode Gel Is Not Ultrasound Gel

The word “gel” does not make contact media interchangeable.

Electrode gel is designed around electrical contact.

Ultrasound gel is designed around acoustic coupling.

Do not substitute ultrasound gel merely because it is water-based, slippery, or visually similar.

Likewise, a conductive electrode gel should not automatically be used with diagnostic or therapeutic ultrasound.

Product selection should follow the energy modality and device instructions.

Are Spectra 360 and SignaGel Sterile?

No.

Spectra 360's current 2026 IFU explicitly identifies it as nonsterile and limits use to external application on intact, unbroken skin.

It is nonsterile and intended for intact skin.

Neither:

  • creates a sterile field;
  • sterilizes skin;
  • should be used on an open wound merely because it is bacteriostatic; or
  • should be substituted for a sterile conductive medium when sterility is required.

Does "bacteriostatic" mean "sterile"?

No.

Both products are described as bacteriostatic, but that characteristic is separate from sterility.

A bacteriostatic formulation is intended to inhibit bacterial growth under its specified conditions.

A sterile product has undergone a validated sterility process and is specifically labeled sterile.

The terms should never be used interchangeably.

Can Salt-Free Gel Cause an Allergic Reaction?

Any contact medium can be inappropriate for a particular patient.

Spectra 360's current IFU specifically warns against use in patients with a known history of paraben sensitivity.

Balego's current SignaGel documentation carries a similar paraben-sensitivity warning.

Therefore, “salt-free” should not be translated into:

hypoallergenic for everyone

or

no allergy risk.

Review the actual product warnings rather than making decisions from salt content alone.

Do Self-Adhesive TENS Pads Need Salt-Free or Saline Gel?

Usually neither.

Modern self-adhesive stimulation electrodes typically include their own conductive hydrogel.

Adding tube gel underneath them can interfere with adhesion or alter the designed electrode interface.

Separate conductive gel is most relevant when using an electrode system that specifically requires it, such as certain:

  • carbon-rubber electrodes;
  • conductive rubber electrodes;
  • reusable non-gelled electrodes; or
  • diagnostic electrodes designed for separate contact medium.

If a self-adhesive hydrogel pad has dried out or no longer adheres reliably, replacement is generally more appropriate than adding an unrelated conductive gel unless its manufacturer provides a specific rehydration procedure.

Spectra 360 vs. SignaGel vs. Redux: Where Does Redux Fit?

Redux is another electrolyte product family, but its positioning is narrower and more cardiology-focused.

Redux Gel is a highly conductive, non-abrasive, water-based electrolyte gel for routine ECG and compatible defibrillation procedures.

ProductMain distinctionStrongest manufacturer-positioned fit
Spectra 360Salt-free + chloride-free + STAY-WETTENS, monitoring, pediatrics, carbon/rubber electrodes
SignaGelMultipurpose saline electrolyte gelECG, EMG, biofeedback, compatible defibrillation
Redux GelCardiology-oriented electrolyte gelRoutine ECG and compatible defibrillation

Which Electrode Gel Should a Clinic Stock?

Some clinics genuinely benefit from stocking both salt-free and saline gel because the products solve different problems.

Rehabilitation / TENS clinic

Spectra 360 is the stronger first choice when reusable conductive rubber or carbon electrodes are routinely used.

Cardiology / diagnostic ECG

SignaGel may be the stronger general-purpose choice, especially if compatible defibrillation workflows also need a separately applied conductive medium.

EMG / biofeedback practice

SignaGel has the stronger direct Parker indication.

Patient-monitoring department

Spectra 360 deserves particular consideration because Parker specifically emphasizes monitoring, long-duration application, and its STAY-WET formulation.

Multidisciplinary clinic

Stocking both can be more appropriate than forcing one formula into every procedure.

Procedure-Based Selection Guide

Procedure or needBetter starting choiceWhy
TENS with carbon/rubber electrodesSpectra 360Parker specifically recommends this combination.
Long-duration electrode contactSpectra 360Salt-free + STAY-WET positioning
Patient monitoringSpectra 360Explicit manufacturer indication
Pediatric electromedical contactSpectra 360Explicitly identified
Routine ECGEither, depending on the electrode/deviceBoth include ECG.
EMGSignaGelDirect manufacturer recommendation
BiofeedbackSignaGelDirect manufacturer recommendation
Defibrillation requiring separate gelSignaGelSpectra 360 is contraindicated.
Salt-free requirementSpectra 360Salt- and chloride-free
Adhesive conductive gel requiredNeitherConsider Tensive.
Self-adhesive hydrogel TENS padUsually neitherConductive gel already incorporated

Frequently Asked Questions

What is a saline electrode gel?

A saline electrode gel is an electrically conductive contact medium containing an electrolyte formulation. Parker specifically describes SignaGel as an ideal saline electrode gel intended for multiple electromedical applications.

What is a salt-free electrode gel?

It is an electrically conductive electrode medium formulated without salt. Parker's Spectra 360 is specifically both salt-free and chloride-free.

Does salt-free electrode gel still conduct electricity?

Yes.

Spectra 360 is expressly labeled as an electrically conductive electrode gel. Parker states that it works by wetting the skin and reducing skin resistance.

Is saline electrode gel more conductive?

You should not assume that it is clinically superior solely because it contains electrolytes.

SignaGel is highly conductive, and Spectra 360 is electrically conductive but does not provide a simple universal head-to-head ranking, establishing one as better for all electrodes or procedures.

Is Spectra 360 salt-free?

Yes.

Parker specifically identifies it as salt-free and chloride-free.

Is SignaGel salt-free?

No.

Parker describes SignaGel as a multipurpose saline electrode gel.

Is salt-free gel better for long-term monitoring?

Spectra 360 is specifically positioned for prolonged and long-term applications, and its STAY-WET formulation is intended to maintain moisture without frequent reapplication.

Which gel is better for carbon rubber TENS electrodes?

Spectra 360.

Parker specifically identifies it as the preferred product for conductive rubber and carbon electrodes and lists TENS among its applications.

Which gel is better for EMG?

SignaGel has the clearer manufacturer indication because Parker specifically recommends it for EMG.

Which gel is better for biofeedback?

SignaGel.

Biofeedback is one of Parker's specifically listed SignaGel applications.

Can Spectra 360 be used for ECG?

Yes.

ECG is one of its manufacturer-listed procedures.

Can SignaGel be used for ECG?

Yes.

Parker specifically recommends it for ECG.

Can Spectra 360 be used for defibrillation?

No.

Parker expressly states that Spectra 360 is not indicated for defibrillation.

Can SignaGel be used for defibrillation?

Yes, when the connected defibrillator/electrode instructions permit separately applied conductive gel. Parker includes defibrillation among SignaGel's recommended applications.

Is salt-free gel automatically better for sensitive skin?

No.

Spectra 360 is positioned for sensitive and prolonged applications, but its IFU still contains a warning for patients with known paraben sensitivity.

Does salt-free" mean "hypoallergenic"?

No.

Salt content and allergy potential are different questions.

Does saline electrode gel damage electrodes?

Not inherently.

Do not generalize from Spectra 360's long-term electrode-performance positioning to conclude that every saline gel damages compatible electrodes.

Is Spectra 360 adhesive?

No.

It is a conductive gel. If temporary adhesive hold is specifically needed, consider an approved conductive adhesive medium such as Tensive.

View Tensive Conductive Adhesive Gel

Should I put Spectra 360 on self-adhesive TENS pads?

Usually not.

Self-adhesive hydrogel pads already include a conductive contact layer. Follow the pad manufacturer's instructions.

Is NuPrep a salt-free electrode gel?

NuPrep is not a leave-on electrode gel at all. It is an abrasive skin-preparation product used before compatible monitoring electrodes.

Are Spectra 360 and SignaGel sterile?

No.

Both are nonsterile products intended for appropriate intact-skin electromedical use.

Can I use ultrasound gel instead?

Do not substitute ultrasound gel merely because it looks similar. Ultrasound gel is designed for acoustic coupling; these products are designed for electrical contact.

Bottom Line: Salt-Free and Saline Gels Are Different Tools, Not Different Quality Levels

The most important takeaway is that salt-free and saline electrode gels should not be ranked as “good” and “better.”

They represent different formulation strategies.

Spectra 360 is the better starting choice when the application calls for:

  • salt-free and chloride-free contact;
  • prolonged electrode use;
  • patient monitoring;
  • TENS;
  • pediatric electromagnetic procedures; or
  • reusable conductive rubber/carbon electrodes.

Spectra 360 to wet the skin and reduce skin resistance without a traditional saline formulation.

SignaGel is the stronger starting choice when the workflow calls for:

  • multipurpose saline electrode gel;
  • ECG;
  • EMG;
  • biofeedback; or
  • compatible defibrillation requiring a separate conductive medium.

Parker expressly identifies those as SignaGel applications.

So the correct question is not

“Does electrode gel need salt?”

It is:

“Which conductive interface is specified for my procedure, electrode, contact duration and medical device?”

Shop Spectra 360 Salt-Free Electrode Gel

Shop SignaGel Saline Electrode Gel

Read SignaGel vs. Electrode Gel: Uses & Differences

Read the Carbon Rubber TENS Electrode Gel Guide

Browse Balego Conductive Gels, Lotions & Contact Media

Product-Use Notice

Always follow the electrode manufacturer's and the electromedical-device manufacturer's Instructions for Use. Spectra 360 and SignaGel are nonsterile contact media intended for compatible external applications on intact skin. Spectra 360 is specifically contraindicated for defibrillation. Both product families have formulation-specific precautions, including paraben-sensitivity considerations. Do not add separate electrode gel to a pre-gelled disposable electrode unless its manufacturer instructs you to do so.