MW72147 0.080 inch pin-to-alligator clip adapter

Cardinal Health

Uni-Patch Model 770 Pin-to-Clip Electrode Adapters — 0.080", 10/Bag

Ref: PIN-TO-CLIP-ADAPTOR

Category: Electrotherapy Electrodes

SKU: 770

$99.95

Out of stock

🛡 Authorized Dealer - Authentic Products & Manufacturer Warranty

💲Price Match Guarantee

🚚 FREE & FAST Shipping

◆ Easy, Secure Checkout

💬Questions? Contact Our Product Specialists

Uni-Patch™ 0.080" Pin-to-Clip Electrode Adapters — 10/Bag

The Uni-Patch Model 770 Pin-to-Clip Adapter converts a standard 0.080-inch electrotherapy lead-wire pin into an alligator-style clip for compatible flat-tab stimulating electrodes.

It is designed primarily for use with products such as Uni-Patch Uni-Tab Stimulating Electrodes, which use a thin conductive tab instead of the molded pigtail socket or snap connector found on many reusable TENS and NMES pads.

Insert the lead wire’s standard pin into the adapter, then attach the metal clip to the electrode’s flat tab. The clip teeth help grip the tab securely so that stimulation can pass from the device, through the lead wire and adapter, into the compatible electrode. Current product descriptions identify this connector specifically as a standard 0.080-inch pin-to-alligator-clip adapter for Uni-Tab electrodes.

Each package contains 10 individual red adapters.

Compatibility notice: This is a connector adapter only. It does not make an electrically incompatible device suitable for a particular electrode or treatment. Confirm the device, lead wire, pin diameter, electrode connector, waveform, and clinical application before use.

Product highlights

  • Converts a standard 0.080-inch pin to an alligator-style clip
  • Approximately 2 mm pin compatibility
  • Designed for compatible flat-tab stimulating electrodes
  • Recommended companion accessory for Uni-Tab electrodes
  • Useful with qualifying TENS systems
  • Useful with qualifying NMES or EMS systems
  • Useful with qualifying FES systems
  • Metal clip grips the electrode tab
  • Quick connection and disconnection
  • Red adapter housing
  • Ten individual adapters per bag
  • Uni-Patch Model 770
  • Manufacturer number MW72147
  • Reusable accessory when clean and undamaged
  • Does not alter the device’s electrical output

Model 770, manufacturer reference MW72147, with a 0.080-inch pin-to-clip connection and 10 adapters per bag.

Quick specifications

SpecificationProduct detail
BrandUni-Patch™
Current portfolioCardinal Health
Product typeElectrode connector adapter
Model770
Manufacturer numberMW72147
Input connectionStandard 0.080-inch pin
Approximate pin diameter2.0 mm
Output connectionAlligator-style electrode clip
Primary electrode formatFlat-tab stimulating electrode
Primary companion productUni-Patch Uni-Tab electrodes
ColorRed
Package quantity10 individual adapters
Electrical conversionNone
Changes waveform or intensityNo
ReusableYes, while clean, secure, and undamaged
Direct skin contact intendedNo
Balego categoryLead Wires and Accessories
Current Balego availabilityIn stock

The adapter is listed within our lead-wire and electrode-accessory inventory and is intended to bridge the physical connection between a standard pin lead and a tab-style electrode.

What does the Model 770 adapter do?

The adapter changes the electrode-end connector format.

Before adding the adapterAfter adding Model 770
The lead wire terminates in a standard 0.080-inch pinThe lead wire terminates in an alligator-style clip
The pin cannot grip a flat electrode tab directlyThe clip can attach to a compatible flat tab
Lead is suited to pigtail/socket electrodesLead may be used with compatible tab electrodes
No change to electrical outputStill no change to electrical output

The product is useful when a clinic has an electrotherapy lead wire ending in standard pins but wants to use flat-tab electrodes such as the Uni-Tab 7020, 7022, or 7024.

What the adapter does not do

Model 770 does not:

  • Convert the device-side plug
  • Convert a pin into a snap
  • Convert a banana plug into a pin
  • Change a TENS unit into an NMES or FES device
  • Change waveform, frequency, pulse width, intensity, or polarity
  • Make an unapproved electrode appropriate for a device
  • Replace a damaged lead wire
  • Correct a dry or poorly adhering electrode
  • Connect directly to a pigtail electrode socket
  • Turn a diagnostic ECG clip into a stimulation lead
  • Permit connection directly to skin
  • Provide a connection for needles or dry-needling electrodes

For connector-selection guidance, read our TENS/NMES Electrode Connectors: Pigtail vs. Tab. Our guide explains that tab, pin, pigtail, and snap connectors are physically different and must be matched to the lead-wire system.

What is included?

Package levelQuantity
Individual red pin-to-clip adapters10
Matched red-and-black pairsNot included as pairs
Lead wiresNot included
ElectrodesNot included
Stimulation deviceNot included

Each electrode requires its own connection. Therefore:

Treatment setupAdapters required
One channel using two electrodes2
Two channels using four electrodes4
Four channels using eight electrodes8
Complete bag10 adapters

A bag of 10 can support five two-electrode connections, two complete four-electrode setups with two spares, or another configuration using up to 10 tab electrodes.

Best companion product: Uni-Tab electrodes

Uni-Tab Stimulating Electrodes are thin, self-adhering pads with flat-tab connections. We currently offer three models:

Uni-Tab modelManufacturer numberSizePackage
7020ET880451.25" × 1.5"48
7022ET880252" × 2.25"48
7024ET880204" × 2.25"48

Uni-Tab electrodes are intended for compatible TENS, NMES, and FES systems but cannot connect directly to a conventional male pin. Model 770 creates the required clip connection when the rest of the electrotherapy system is compatible.

Why clinics may use this combination

Clinical needUni-Tab plus Model 770 benefit
Existing standard-pin lead wiresAdapters allow the leads to grip flat tabs
Bulk electrode inventoryUni-Tab electrodes are sold in 48-count packages
Quick pad changesThe clip can be released and moved to a fresh tab electrode
Multiple electrode sizesOne adapter style can support the three Uni-Tab sizes
Thin electrode profileTab pads avoid an attached pigtail wire on the electrode
High-volume treatment roomsReusable adapters may remain with the lead-wire setup

Compatibility checklist before ordering

1. Check the electrode-end lead pin

The adapter requires a standard 0.080-inch pin, approximately 2 mm in diameter.

Our 45-inch TENS/NMES replacement lead wires terminate in standard 2 mm electrode pins. They will provide the appropriate electrode-side connection for Model 770, but the lead wire’s 2.35 mm device-side plug must also fit the stimulator.

2. Check the electrode

The electrode must have a compatible flat conductive tab that can be gripped securely by the adapter clip.

Do not assume compatibility with:

  • Pigtail sockets
  • Snap electrodes
  • Diagnostic ECG tabs
  • Proprietary adhesive pads
  • Wireless TENS pads
  • Magnetic connectors
  • EMPI safety connectors
  • Needle electrodes
  • Conductive garments

3. Check the device-side lead connection

The adapter is installed at the electrode end of a lead wire. It does not change the plug that connects the lead to the stimulator.

A lead may have the correct 2 mm electrode pin but still have the wrong plug for the device.

4. Check the modality

Confirm that the device, electrode, and lead-wire system are appropriate for the selected modality, such as TENS, NMES, EMS, FES, or another device-approved stimulation mode.

5. Check how many adapters are required

Count one adapter for each tab electrode used simultaneously.

Connection example

A compatible setup may follow this sequence:

Stimulator → compatible device-side lead plug → lead wire → 0.080-inch pin → Model 770 adapter → alligator clip → Uni-Tab electrode

Every part of that chain must be physically and electrically compatible.

How to connect the adapter

Always turn off the stimulation device and reduce all channels to zero before making or changing connections.

1. Inspect the equipment

Do not use the adapter if:

  • The metal is bent or corroded.
  • The clip will not open or close correctly.
  • The pin receptacle is loose.
  • The insulation or red housing is cracked.
  • The clip's teeth are damaged.
  • Output cuts in and out when the connector is moved.

Also inspect the lead wire and electrode.

2. Insert the electrode pin

Push the standard 0.080-inch male lead-wire pin fully into the adapter’s receiving end.

The connection should be secure without forcing, bending, or twisting the pin.

3. Open the clip

Compress the adapter mechanism to open the alligator-style jaws.

4. Attach the electrode tab

Position the clip on the compatible conductive tab. The jaws should grip the tab securely without contacting the patient’s skin.

Do not attach the clip to:

  • The adhesive gel area
  • The patient’s skin
  • Jewelry
  • A lead wire
  • A damaged or torn portion of the electrode
  • Another clip

5. Apply the electrode

Place the electrode on clean, dry, intact skin according to the device instructions and treatment plan.

Make sure the clip and lead wire do not pull the electrode from the skin.

6. Begin at zero intensity

Turn the device on and increase output gradually.

Stop and inspect the setup if stimulation is

  • Intermittent
  • Sharp
  • Burning
  • Concentrated at one edge
  • Absent from one electrode
  • Significantly different from the expected sensation

7. Disconnect safely

Return all channels to zero and turn off the device.

Release the clip from the electrode tab before removing or storing the equipment. Do not pull the lead wire or adapter away by force.

Adapter comparison

Which Uni-Patch adapter do you need?

AdapterConversionBest useBalego link
Model 770 / MW721470.080" pin to alligator clipConnecting standard pin leads to flat-tab electrodesPin-to-Clip Adapter
Model 745 / MW722650.080" pin to snapConnecting standard pin leads to compatible snap electrodesPin-to-Snap Adapter
Model 721B / MW72160Banana plug to 0.080" pinConverting a compatible banana connection to a standard electrode pinBanana-to-Pin Adapter
No adapterDirect 2 mm pin into pigtail socketStandard reusable hydrogel pads with pin-compatible leadsBalego Electrodes

The Model 745 has a 0.080-inch pin-to-snap adapter, and the Model 721B has a banana-plug-to-0.080-inch-pin adapter. Both are sold in bags of 10. (Balego)

Choose Model 770 when:

  • Your lead wire ends in a standard 0.080-inch pin.
  • Your electrode has a flat conductive tab.
  • You are using compatible Uni-Tab electrodes.
  • You need a reusable clip connection.
  • You want to retain an existing compatible pin-style lead-wire system.

Choose Model 745 when:

  • Your lead wire ends in a standard pin.
  • Your electrode has a metal snap.
  • A secure snap connection is required.
  • The device and electrode instructions permit the combination.

Choose Model 721B when:

  • The starting connector is a compatible banana plug.
  • The electrode requires a conventional pin connection.
  • You are adapting a clinical stimulation lead system rather than attaching to a tab.

Choose no adapter when:

  • The electrode already has a pigtail socket.
  • The lead wire’s 2 mm pin fits directly.
  • Reducing the number of connection points is preferable.

Model 770 vs. using standard pigtail electrodes

FeatureModel 770 plus Uni-TabStandard pigtail electrode
Electrode connectionClip grips flat tabPin inserts into molded socket
Additional adapterYesNo
Electrode package optionsUni-Tab 48-packOften 4, 16, 40, or case quantities
Electrode profileThin tab-style constructionAttached pigtail connection
Setup speedQuick once adapters are installedDirect connection
Connection pointsMoreFewer
Best fitClinics already using tab electrodesMost portable home TENS and NMES systems
ExamplesUni-Tab ElectrodesBalego TENS/NMES/FES Electrodes

A direct pigtail pad is generally simpler when the user does not specifically need tab electrodes. The adapter-and-tab combination is most useful for clinics that prefer Uni-Tab’s thin profile, bulk packaging, sizes, or quick clip connection.

Troubleshooting guide

ProblemPossible causeRecommended action
No stimulationThe lead pin is not fully insertedTurn off device and reconnect securely
No stimulationThe clip is not gripping the conductive tabReposition the clip
One channel does not workDamaged lead, adapter, or electrodeSwap known-good components one at a time
Output cuts in and outLoose pin or worn clipReplace the affected adapter or lead
The clip slips from the tab.The clip is not fully engaged or tab is damagedReattach or replace the electrode
The clip touches the skinThe adapter is positioned too close to the electrode bodyTurn off device and reposition it.
Electrode liftsLead-wire tension is pulling the padSupport or reroute the cable
Stimulation feels sharpThe electrode is lifting, dry, or contaminatedReplace or reposition the electrode
The adapter does not fit the pin.The pin is not a standard 0.080-inch sizeDo not force it; verify the lead specification
The adapter fits pin but not the device.Wrong device-side lead plugSelect the correct complete lead wire
The clip becomes hotLoose or damaged connectionStop treatment immediately and replace the damaged component
Red adapters cause channel confusionAll adapters look the sameLabel the leads and follow the device’s channel markings

Care and storage

After use:

  1. Turn off the device.
  2. Release the clip from the electrode tab.
  3. Disconnect the adapter by holding its housing rather than pulling the lead wire.
  4. Inspect the clip and pin receptacle.
  5. Remove dry gel or debris according to the product and facility instructions.
  6. Allow the adapter to dry completely before storage.
  7. Store loosely with the lead wires.
  8. Keep away from moisture, corrosive cleaners, extreme heat, and sharp objects.

Do not:

  • Immerse the adapter unless the manufacturer’s instructions expressly permit it.
  • Use it with damaged insulation.
  • Bend the pin to make it fit.
  • Force an oversized pin into the adapter.
  • Clip it directly to skin.
  • Permit the metal clip to contact another adapter during active stimulation.
  • Use it for ECG acquisition unless the specific system expressly permits it.
  • Use it to connect electrical stimulation to needles.

Reviews

Rating Snapshot

5 Stars
0
4 Stars
0
3 Stars
0
2 Stars
0
1 Stars
0

Overall Rating

0 Reviews

Review this Product

Sign in to write a review.

No reviews yet — be the first to share your thoughts.

FAQs

What is a pin-to-clip electrode adapter?

It is a connector that accepts a standard electrode pin and provides an alligator-style clip for attaching to a compatible flat-tab electrode.

What size pin does Model 770 accept?

It is designed for a 0.080-inch pin, approximately 2 mm in diameter.

What electrodes is it designed for?

It is primarily designed for compatible flat-tab stimulating electrodes, including Uni-Patch Uni-Tab Electrodes.

How many adapters are included?

The Balego package contains 10 individual red adapters.

Are there five red-and-black pairs?

No. The listing describes 10 red adapters, not five matched color pairs.

How many adapters do I need for a dual-channel treatment?

A typical dual-channel setup using four electrodes requires four adapters—one for every electrode.

Will this adapter fit every TENS lead wire?

No. The electrode-end lead pin must be the correct 0.080-inch size. The opposite end of the lead wire must also fit the stimulation device.

Will it work with Balego’s 45-inch lead wires?

Those lead wires terminate in standard 2 mm electrode pins, so the electrode-side connection may match. Their 2.35 mm device-side plug must also fit the stimulator.

Does it convert a pin to a snap?

No. For snap-style electrodes, compare the Uni-Patch Model 745 Pin-to-Snap Adapter.

Does it convert a banana plug to an electrode pin?
Does the adapter change treatment intensity?

No. It is a passive physical connector. It does not intentionally change intensity, waveform, frequency, pulse width, or treatment mode.

Is it used for TENS?

It can be used in a compatible TENS setup when the lead wire, adapter, electrode, and device instructions all match.

Can it be used for NMES or EMS?

Yes, when the complete NMES or EMS setup supports the connector and selected electrode.

Can it be used for FES?

It may be used in a compatible clinician-directed FES setup. Connector compatibility does not by itself establish that the electrode size or device is appropriate.

Can it be used with diagnostic ECG tab electrodes?

Do not assume so. ECG tab electrodes are designed to receive cardiac signals, while Uni-Tab electrodes deliver stimulation. The equipment, lead wires, connector construction, and intended use are different.

Can it clip directly to a carbon-rubber electrode?

Only if the carbon electrode is specifically designed with an appropriate tab connection. Do not attach the clip arbitrarily to an electrode surface.

Does red indicate positive polarity?

The adapter housing is red, but users should rely on the stimulation device, lead-wire markings, channel identification, and treatment instructions—not the adapter color alone.

Can it be reused?

Yes, while it remains clean, secure, corrosion-free, and physically undamaged. Replace it when the connection becomes loose or intermittent.

Why is the adapter connection intermittent?

Common causes include:

  • An undersized or damaged lead pin
  • Incomplete insertion
  • Worn clip tension
  • Corrosion or contamination
  • A torn electrode tab
  • A damaged lead wire
  • Lead-wire tension pulling on the adapter
Can it be used in water?

No. Do not operate electrotherapy equipment or its connectors in a bath, shower, pool, or another wet environment.